50 stillman road, north haven ct 06473 call 800.243.2000 or 203.239.2090

Splitex Plate Installation And Calibration

Instructions for Recalibrating:
CN, CT, CP, CR, Artex Articulators and Carbon-Series Mounting Unit click here

Instructions for application of Loctite® 330
click here

Instructions For Calibrating:
CN, CT, CP, CR, Artex Articulators and Carbon-Series Mounting Unit

Not for use with the Artex Classic Series

You will need:

01. Magnet Removal tool
02. Assembly Tool
03. Mounting Plate
04. Magnet Pot
05. Mounting Screws
06. Base Plate – Maxilla
07. Adjustment Plate – Mandible
08. Adhesive Plate (note beveled edge)

Petroleum jelly
Loctite 330® adhesive with activator
+10mm Spacer Plate*

NOTE:* If calibrating with a “Classic Artex” 110 mm key, attach a “Spacer Plate, +10 mm”, (46011, not pictured) to the upper Splitex Plate now (magnet to magnet.)

 

The Steps for Calibrating:

Step 1:

Set the articulator to average-value settings*
(standard working conditions).

* 30-35° condylar inclination * 10-15° Bennett angle * Anterior guidance pin at 0

 

Step 2:

Remove blue model plates from upper and lower arms

 

Step 3:

Remove upper magnet with magnet removal tool

 

Step 4:

Loosen and unscrew magnet pot with assembly tool

 

Step 5:

Magnet pot removed

 

step 6:

Screw in the mounting plate all the way, using the assembly tool

 


step 7:

Mounting plate inserted

 

Step 8:

Unscrew the magnet pot from upper magnetic calibration plate. Place that upper plate onto the upper arm of the articulator

 

Step 9:

Attach upper magnetic calibration plate to the upper arm of the articulator using the socket-head screw

 

Step 10:

Screw in the magnet pot to the upper magnetic claibration plate

 

Step 11:

Magnet pot inserted.

 

Step 12:

Replace the magnet.

 

Step 13:

Magnet replaced

 

Step 14:

Remove lower magnet

 

Step 15:

Loosen and unscrew magnet pot with assembly key

 

Step 16:

Magnet pot removed

 

Step 17:

Screw mounting plate all the way in, using assembly tool

 

Step 18:

Mounting plate inserted

 

Step 19:

Isolate the Carbon surface in the area of the adhesive plate with a thin layer of petroleum jelly

 

Step 20:

Set in the adhesive plate, with chamfering to the rear (Important for later removal!)

 

Step 21:

Attach the adhesive plate with socket-head screw

 

Step 22:

Turn articulator upside down. Open the centric locks. Note: if calibrating to Classic artex, now place the 10 mm spacer plate on the magnetic calibration plate. (For use on UPPER ARM only. Do not place spacer on lower arm)

 

Step 23:

Place Splitex Key (C6010) on the upper magnetic calibration plate (or when calibrating to classic Artex, use key 46010, and place it on the 10 mm spacer plate)

 

Step 24:

TOP to TOP

 

Step 25:

Place the lower magnetic calibration plate (with the magnet) on the key.

 

Step 26:

Clean the adhesive plate, and the lower (milled) side of the lower magnetic calibration plate til grease-free

 

Step 27:

Spray the activator on the adhesive plate

 

Step 28:

Put on two rows of adhesive of glue to the left and right of the drilled areas, not into the drilled areas!

 

Step 29:

Close the articulator

 

Step 30:

Close the centric locks

 

Step 31:

Press the articulator front for approximately 5 seconds gently against the incisal pin. Let the adhesive set for 10 minutes or until completely set. Do not move the articulator in the meantime!


 

Steps for Recalibrating:

Step 1:

Remove the magnet

 

Step 2:

Unscrew the magnet pot

 

Step 3:

Unscrew the socket-head screw

Place a plaster knife between the chamfering of the glue plate and the splitex magnetic calibration plate and tilt it to separate!

 

Instructions for application of Loctite® 330:

1. Clean the adhesive plate and bottom (milling holes) of the mandibular plate to be sure it is free of grease.

2. Use Loctite adhesive to glue the mandibular plate to the adhesive plate. Spray activator onto the bottom Splitex plate for the mandible. Apply two parallel lines of adhesive centered between the edges and the center milling holes of the adhesive plate. Immediately close the articulator and slightly press on the anterior pin for 5 seconds until the adhesive has completely adapted on both sides. Let set for 10 minutes. The calibration procedure is finished when the adhesive has completely set.

Mounting Unit Exception: The one-piece lower Splitex plate for the Mounting Unit (C6020) is designed to attach solely by gluing, not with a screw.

Important Notes: For cleaning and care of Splitex plates, use only cleaning agents that are suitable for light metals, following the manufacturer’s instructions. Do not clean Splitex plates in an ultrasonic bath, as the anodized coating may be damaged. Do not use cleaning agents with a pH of <5 or >8 since these might damage the anodized surfaces.

Since heat will irreversibly reduce the power of the magnets, do not steam-clean or boil the magnets.